《高绝缘电阻测量仪(高阻计)校准规范试验报告2.docx》由会员分享,可在线阅读,更多相关《高绝缘电阻测量仪(高阻计)校准规范试验报告2.docx(5页珍藏版)》请在taowenge.com淘文阁网|工程机械CAD图纸|机械工程制图|CAD装配图下载|SolidWorks_CaTia_CAD_UG_PROE_设计图分享下载上搜索。
1、高绝缘电阻测量仪(高阻计)校准规范试验报告标准器不同电压测量试验、样机的基本信息计量器具名称型号、规格、准确度样机编号兆欧表检定装置ZX119-10/10kV1214182二、试验所用设备序号仪器设备名称规格(型号)准确度编号12数字式高阻检 定仪静电计TY13016517A电阻:100Q-10T QUrel= (3 X 101 X10-2),仁2; MPE: 0.15%0011124063三、试验地点:中国计量科学研究院和平里院区2号楼120B室 四、试验时间:2022年07月11日至2022年07月12日五、试验人员:梁波六、型式评价环境条件:温度:20.0至I 20.5湿度:40%RH
2、到 43%RH七、试验结果JJF XXX XX XXX1 .测量盘100GQ标称值/G。测量电压/V实际值/ GQ标称值/GQ测量电压/V实际值/ GQ1001099.24100010965100100100.101000100997.7100250100.1610002501000.2100500100.1610005001001.41001000100.17100010001002.31001000100.0100010001000.31002500100.1100025001001.31005000100.0100050001001.81001000099.91000100001002.
3、02 .测量盘10GQ标称值/GQ测量电压/V实际值/ GQ标称值/GQ测量电压/V实际值/ GQ101010.0091001099.641010010.008100100100.151025010.029100250100.181050010.028100500100.1910100010.0271001000100.1810100010.0271001000100.2110250010.0281002500100.2410500010.0281005000100.25101000010.02610010000100.203 .测量盘1GQ标称值/GQ测量电压/V实际值/ GQ标称值/GQ测
4、量电压/V实际值/ GQ1101.0016101010.0011001.00161010010.0312501.00001025010.0315001.00061050010.03110001.000910100010.03110001.001110100010.026125001.001110250010.030150001.001110500010.0301100001.0009101000010.0254 .测量盘100M。标称值/MQ测量电压/V实际值/ MQ标称值/MQ测量电压/V实际值/MQ10010100.021000101002.6100100100.2110001001002
5、.7100250100.2610002501002.2100500100.2510005001002.41001000100.24100010001002.41001000100.25100010001001.91002500100.24100025001001.41005000100.24100050001001.710010000100.21100()100001001.75 .测量盘10MQ标称值/MQ测量电压/V实际值/ MQ标称值/MQ测量电压/V实际值/M。10109.9891001099.99101009.99310010099.94102509.99510025099.9510
6、5009.99410050099.941010009.993100100099.941010009.993100100099.881025009.993100250099.931050009.992100500099.956 .测量盘1M。标称值/MQ测量电压/V实际值/ MQ标称值/M。测量电压/V实际值/M。1100.999810109.99111000.9999101009.99512500.9987102509.99715000.9993105009.996110000.99961010009.991110000.99981010009.991125000.99981025009.995八、试验结论:根据试验结果可知该兆欧表检定仪在2500Y5000VJ0000V下的测量结JJF XXX XX XXX果与100V的测量结果1OOG。最大差0.2%, 1000G。最大差0.43%,可以满 足最大允许误差在2%的高阻计校准。其余电阻值在2500V,5000V,10000V下 的测量结果与100V的测量结果最大差小于0.1%,可以满足最大允许误差1% 的高阻计校准。可以使用该兆欧表检定仪的额定电压下的电阻测量结果对高 阻计100V进行校准。